TOCII CA72TOC Identification
Endress+Hauser 7
2 Identification
2.1 Nameplate
Compare the order code indicated on the nameplate with the ordering structure and your order.
The nameplate bears the following information:
Manufacturer data
Order code (device version)
Serial number
•Measuring range
Outputs and communication
Power supply
Degree of protection
(Permitted) ambient conditions
2.1.1 Product structure
Measuring range, adjustable
A 0.25 to 600 mg/l TOC
B 1 to 2400 mg/l TOC
C 2.5 to 6000 mg/l TOC
D 5 to 12000 mg/l TOC
Y Special version according to customer specification
Sample transfer
1 1 measuring point
2 2 measuring points; time controlled
3 2 measuring points; external controlled
Operation mode
A Batch mode
Power supply
0 230 VAC, 50 Hz
1 115 VAC, 60 Hz
2 115 VAC, 50 Hz
3 230 VAC, 60 Hz
9 Special version according to customer specification
Sample preparation
A 1 x PA-2 PVC, 1 to 8 m
3
/h (4.4 to 35 gpm) wastewater (not with
CA72TOC-*2*********)
B 1 x PA-3 PVC, 0.1 to 1.0 m
3
/h (0.4 to 4.4 gpm) wastewater (not with
CA72TOC-*2*********)
C 1 x PA-9 PP, 0.1 to 1.0 m
3
/h (0.4 to 4.4 gpm) wastewater (not with
CA72TOC-*2*********)
D 2 x PA-2 PVC, 1 to 8 m
3
/h (4.4 to 35 gpm) wastewater (not with
CA72TOC-*1*********)
E 2 x PA-3 PVC, 0.1 to 1.0 m
3
/h (0.4 to 4.4 gpm) wastewater (not with
CA72TOC-*1*********)
F 2 x PA-9 PP, 0.1 to 1.0 m
3
/h (0.4 to 4.4 gpm) wastewater (not with
CA72TOC-*1*********)
Y Special version according to customer specification