Commissioning TOCII CA72TOC
38 Endress+Hauser
6.3.5 Menu SETTING - SET DEFAULTS
6.3.6 Menu SETTING - MEASURING SITE
You can give the measuring point a name.
a0012582-en
Fig. 18: "Programming" menu
In the factory setting, the name is MEASURING SITE.
• Select the position of the character using the cursor buttons T and U.
• Select the character using the V and W cursor buttons.
• Pressing the "1" key switches the character to "A".
6.3.7 Menu LISTS - MIN MAX AVERAGE
a0012585-en
Fig. 19: "Lists" menu
This menu item displays the maximum, minimum and average measured values for the days saved.
Step Display Description
1
2 • If you are sure that you want to reset to
the factory settings, press the [E] key.
• If you want to keep the current values,
then press the operation key.
SETTING
LISTS
PROGRAMMING
TEST
RANGE DATA
SET CLOCK
ALARM LIMITS
BASIC DATA
MEASURING SITE
SET DEFAULTS
ARE YOU SURE?
SET DEFAULTS
[E] TO CONFIRM
SETTING
LISTS
PROGRAMMING
TEST
RANGE DATA
SET CLOCK
ALARM LIMITS
BASIC DATA
SET DEFAULTS
MEASURING SITE
SETTING
LISTS
PROGRAMMING
TEST
MAX MIN AVERAGE
MAINTENANCE RECORD
ALARM RECORD
SHOW COMPLETE RECORDS