TOCII CA72TOC Maintenance
Endress+Hauser 95
7.9 Cleaning
7.9.1 Screen flushing
During a screen flush, the sample conditioning system and the filter used are flushed with
pressurized water. This takes place via solenoid valve MV5.
a0012726
Fig. 52: Position of the screen
Screen flushing runs automatically after startup, and can be started in three ways:
Activated manually
Activated remotely
Activated automatically
Operation starts automatically on completion of the screen flush process.
Step Display Description
1
PUMPS
CALIBRATION
SERVICE
CLEANING
FILTERS
SCREEN FLUSH
POWER-FLUSH
BYPASS-SCREEN
STRIPPER+SEPARATION
OPEN GAS CICUIT
COMBUSTION PIPE
LEAKAGE TEST