TOCII CA72TOC Commissioning
Endress+Hauser 45
6.3.13 Menu TEST - TEST FREQUENCY I/O
a0012616-en
Fig. 25: "Test" menu
Test the inputs and outputs to test the function of the I/O card.
6.3.14 Menu TEST - TEST COM2
a0012617-en
Fig. 26: "Test" menu
Displays the transmission data for the computer interface RS 232.
The menu item allows users to test the transmission of data with an external terminal.
Once the data connection has been established, the line "abcdef-ABCDEF-01234567890" can be
sent to the terminal by pressing a key, e.g. the [0] key.
Keystrokes at the external terminal are shown on the display.
Press the [E] key to exit the menu item.
Output Description Switch status OFF Switch status ON
SA1 Changeover between standard and sample Sample Standard solution
SA2 Flushing valve for power flushing Strip chamber flushing
switched off
Strip chamber flushing
switched on
SA3 O
2
supply, tube furnace controller, Peltier
cooler controller, membrane compressor
Consumer load switched off Switch status during
measuring operation
SA4 Changeover between standard 1 and
standard 2
Standard 1 Standard 2
SA5 Screen flushing valve Screen flushing valve off Screen flushing valve on
SA6 Optional: Changeover between channel 1
and channel 2
Channel 1 Channel 2
SA7 Carrier gas flushing valve Carrier gas flushing off Carrier gas flushing on
SA8 Dosing valve Dosing valve open Dosing valve closed