TOCII CA72TOC Maintenance
Endress+Hauser 55
7 Maintenance
Maintenance work has to be performed at regular intervals to ensure the efficient operation of the
analyzer.
#
Warning!
Before attempting to service the device, you must be completely familiar with the maintenance
processes and have perfectly understood the tasks involved. All the maintenance tasks described in
this chapter should only be carried out by suitably trained specialists. Mistakes made during
maintenance can result in inaccurate operation and pose a safety hazard.
The maintenance intervals depend greatly on your application. For this reason, adapt the
maintenance intervals to your specific requirements but make sure to perform these maintenance
tasks regularly!
Interval Maintenance task
At least once per week Visual inspection
Check sample conditioning (see corresponding Operating Instructions)
At least once per month Check feed rate of pump P1/P4 and P2
Replace standard
Every 3 months at the latest Clean strip and separation chamber
Replace glass bead
Adjust pH sensor
Replace pump hoses
If salt quantity > 1 g/l, then
perform the following every 3
months at the latest:
Replace acid filter
Replace catalyst
Clean combustion pipe