TOCII CA72TOC Commissioning
Endress+Hauser 35
6.3.1 Menu SETTING - RANGE DATA
a0012553-en
Fig. 15: Menu "Setting"
SETTING
LISTS
PROGRAMMING
TEST
RANGE DATA
SET CLOCK
ALARM LIMITS
BASIC DATA
SET DEFAULTS
MEASURING SITE
Parameter Unit Factory
setting
Description
SCALE mg TOC/l 1000 Use this function to enter the maximum concentration for
your measuring point here. This value determines the scale
end value for the graphic interface
Enter two independent values for the two-channel version.
SCREEN FLUSH n/day 0 The number of automatic flush cycles for the bypass screen
per day (recommended value: 2).
POWER FLUSH n/day 0 The number of automatic power flush cycles for the
stripping vessel and separation chamber per day
(recommended value: 2).
DAYBREAK 0 Time of day changeover.
This parameter controls automatic events; e.g. automatic
calibration starts 90 minutes before the day changes (i.e.
5.00 a.m. if the day changes at 6.30 a.m.).
6.5 is equivalent to 6.30 a.m.
PAUSE CYCLE s 0 Interval between 2 measurements.
P1 (B) ml/min 7.5 Feed rate of pump P1.
P4 (B) ml/min 6.0 Feed rate of pump P4 (optional).
VOL.BATCH μl 300 Dosing volume for a batch. Increasing the volume increases
the sensitivity of the measuring system but also increases
the salt load.
STANDARD C1 mg/l 80 Concentration of standard solution C1.
STANDARD C2 mg/l 720 Concentration of standard solution C2.
CAL./ADJUSTM. n
DAY
3 Here you can specify after how many days a calibration or
adjustment should be performed. The automatic function is
switched off with the value 0.
CAL/ADJUSTMENT 2 Use this function to specify which function should be
executed.
1 - Calibration
•2 - Adjustment
The function is executed 90 minutes before the day
changes.