Maintenance TOCII CA72TOC
64 Endress+Hauser
a0012364
Fig. 34: Pump P1
Pump P1: Sample hose to strip chamber has an ID of 2.79 mm (0.11") and the color code
violet-white (VT-WH). The hose to extract condensate has an ID of 0.76 mm (0.03")and the color
code black-black (BK-BK).
Pump P4 (only for version "with predilution"): Sample hose to static mixer has an ID of 2.79 mm
(0.11") and the color code (VT-WH).
1
2
3
4
5
6
7
8
1 Pump retainer 5 Motor
2 Pump head 6 Condensate hose (BK-BK) 0.76 mm (0.03") ID
3 Hose cassette (sample hose) 7 Sample hose (VT-WH) 2.79 mm (0.11") ID
4 Hose cassette (condensate hose) 8 Adjusting screw
Step Display Description
1
2
#
Warning!
There is a risk of infection when working
with wastewater! For this reason, wear
protective goggles, protective gloves and
protective clothing.
Press the [E] key.
The strip and separation chambers are
flushed with pressurized water.
PUMPS
CALIBRATION
SERVICE
CLEANING
FILTERS
P 1/4 REPLACE HOSE
PUMP 2 REPLACE HOSE
PUMP 3 REPLACE HOSE
CALIBRATE PUMP 2
REPLACE HOSE P1/4
CAUTION! LIQUID CAN LEAK OUT
WEAR PROTECTIVE CLOTHES
[E] TO CONFIRM