TOCII CA72TOC Maintenance
Endress+Hauser 67
You can enter the new value after 60
seconds.
The value typically lies between 5.5 and 7
ml (0.18 and 0.24 fl.oz).
The amount entered is displayed so you
can check it is correct.
Press the [E] key.
8 1. Seal the strip chamber inlet with a
separate plug (e.g. plug for leak test).
2. Connect hose P4 at the pressure side.
3. Press the [E] key.
Step Display Description
REPLACE HOSE P1/4
[E] TO CONFIRM
PREVIOUS VOLUME DELIVERED x.x ml
ENTER ml?
REPLACE HOSE P1/4
[E] TO CONFIRM
[>] START/STOP PUMP 4 (60sec)
NEW VOLUME DELIVERED x.x ml
REPLACE HOSE P1/4
[E] TO CONFIRM
CONNECT HOSE P4
SHUT STRIP CHAMBER INLET WITH
SEPARATE PLUG