61 / 80
8 • Status Messages, Error Messages
80113-401_V24023
2023-05
An Error has been detected.
Activation not possible.
The waste gas tube is obstructed. Check whether the waste gas tube is
kinked or blocked.
Unit malfunction If the error continues to be displayed,
inform Technical Service.
An Error has been detected.
Please use another probe.
Probe malfunction Connect a different probe.
Unit malfunction If the error continues to be displayed,
inform Technical Service.
Footswitch is not compati-
ble. Activation not possiblel.
The software version of the footswitch is not
suitable.
Connect a footswitch with a suitable
software version.
You have connected a footswitch for a different
type of unit.
Connect the CRYO footswitch (white
pedal).
Unit malfunction If the error continues to be displayed,
inform Technical Service.
Unit is overheated. Activa-
tion not possible.
The ambient temperature at which the unit was
stored was too high.
Allow the unit to acclimatize at room
temperature for approx. 3 hours.
The ambient temperature during operation is too
high.
Operate the unit at a room temperature
of between +15 °C - +40 °C.
There is poor ventilation around the unit housing. Install the unit in such a way that there is
an unobstructed circulation of air around
the housing. Installation in confined wall
recesses is prohibited.
Unit malfunction If the error continues to be displayed,
inform Technical Service.
Unit is too cold. Activation
not possible.
The ambient temperature at which the unit was
stored was too low.
Allow the unit to acclimatize at room
temperature for approx. 3 hours.
The ambient temperature during operation is too
low.
Operate the unit at a room temperature
of between +15 °C - +40 °C.
Unit malfunction If the error continues to be displayed,
inform Technical Service.
Low input pressure. Please
check gas supply.
The gas bottle is almost empty. Connect a new gas bottle.
The bottle valve is not fully open. Open the bottle valve completely.
The gas bottle is too cold. Allow the gas bottle to acclimatize at
room temperature for approx. 8 hours.
Unit malfunction If the error continues to be displayed,
inform Technical Service.
High pressure at waste gas
nozzle. Please check waste
gas nozzle.
The waste gas tube is obstructed. Check whether the waste gas tube is
kinked or blocked.
Unit malfunction If the error continues to be displayed,
inform Technical Service.
High input pressure. Please
check gas supply.
The gas bottle is too full or too warm. Allow the gas bottle to acclimatize at
room temperature for approx. 8 hours.
Unit malfunction If the error continues to be displayed,
inform Technical Service.
Error text Possible cause(s) What should you do?