METALJET D2+ TRANSPORT AND STORAGE
Revision 19, November 2016 Page 3-2
Figure 3-2. Pay special attention to the cooling-water connections and the jet-exit port when placing the
electron gun/x-ray head on a table. They may be damaged unless positioned slightly outside the edge of the
table.
3.2 Storage
The MetalJet D2+ source should only be stored in an upright position.
Check that all openings on the MetalJet D2+ source are properly closed in order to prevent dust, dirt,
or other contaminants from entering the system during storage.
Store the MetalJet D2+ source in the temperature range 20-40°C (68-104°F).
Store the MetalJet D2+ source in a dry place.
- If it cannot be stored in a dry environment it (as a whole unit or divided into its sub-assembly
units) must be airproof shrink-wrapped in a plastic bag together with some bags of desiccant.