METALJET D2+ TECHNICAL DATA
Revision 19, November 2016 Page 11-2
Parameter
Value
Cooling water specification
Water specification
According to water cooling system instructions
Min pump flow rate capacity
3 l/min
Min pump pressure capacity
5 bar
Typical water flow rate
4 l/min at ΔP = 3 bar
Water temperature
Alloy dependent, typically 30 ± 0.2°C (86 ± 0.4°F)
Water connection
3/8” hose barb
Cooling capacity
300 W minimum
Environmental operating conditions
Installation location
Weather protected (indoors)
Installation altitude
2000 m max.
Temperature
Any temperature in the range 15-30°C (59-86°F), but
preferably stable within ±0.5°C (±0.9°F)
Degree of pollution
2
Relative humidity of air
Max 85% (no condensation)
Overvoltage category
II
Mains supply voltage fluctuations
Up to ±10% of the nominal voltage
Storage conditions
Room temperature
20-40°C (68-104°F)
Relative humidity of air
Max 85% (no condensation)