16-3
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
(b) Select [GAIN];
(c) Select [×0.25], [×0.5], 1] or [×2].
16.4.2 Setting theApneaAlarmTime
(a) Select RESP parameter area and the [RESP SETUP]menu will be displayed;
(b) Select [APNEA ALM]and then select [10S], [15S], [20S], [25S], [30S], [35S], [40S]or [OFF].
[10S] meansthat if apnea lasts for more than 10s and RESP alarm function is enabled, the
monitor will give an apnea alarm.
[OFF] means the apnea alarm function is disabled.
16.4.3 Setting the RESP Lead
RESP lead is where the current respiratory waveform comes from. There are RA-LA (I) and RA-LL (II) to
choose.
(a) Select RESP parameter area and he [RESP SETUP]menu will be displayed;
(b) Select [RESP LEAD] and select [RA-LA (I)]or[RA-LL (II)].
16.4.4 Setting the RESP Filter
Enhancing filter is to enhance filtering the interference of heart beat.
(a) Select RESP waveform area and he [RESP WAVE]menu will be displayed;
(b) Select [FILTER] and select [ON] or [OFF] to enable or disable the RESP filter function respectively.