EasyManuals Logo

Fazzini FM5000C User Manual

Fazzini FM5000C
162 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
18
-
7
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
(7) Obese Patients
A thick layer of fat around a limb damps oscillations from the artery, thus preventing them from
reaching the cuff. The accuracy is lower than the normal one.
18.5 NIBP Settings
18.5.1 Setting the Measure Mode
Measure mode includes adult, child and neonate, which is the same as the patient type.
(a) Select NIBP parameter area and the [NIBP SETUP]menuwill be displayed;
(b) Select [MEASURE MODE]
(c) Select[ADU], [PED] or[NEO] ;
18.5.2 Setting the Interval Time
(a) Select NIBP parameter area and the [NIBP SETUP]menuwill be displayed;
(b) Select [INTERVAL];
(c) Select [MANUAL] to conduct a manual measurement;
(d) Select [1MIN], [2MIN], [3MIN], [4MIN], [5MIN], [10MIN], [15MIN], [30MIN], [60MIN], [90MIN],
[120MIN], [180MIN], [240MIN] or [480MIN] for auto measurement;
18.5.3 Setting the Pressure Unit
(a) Select NIBP parameter area and the [NIBP SETUP] menuwill be displayed;
(b) Select [UNIT];
(c) Select [mmHg] or [kPa];
18.5.4 Setting the Pre-inflation Value
(a) Select NIBP parameter area and the [NIBP SETUP] menuwill be displayed;
(b) Select [INFLATION] and select the suitable pre-inflation value;
Neonates: the range of the pre-inflation value is 60-120 mmHg, the default value is 100mmHg
Pediatric: the range of the pre-inflation value is 80-200 mmHg, the default value is 120mmHg
Adults: the range of the pre-inflation value is 80-240 mmHg, the default value is 160mmHg;

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fazzini FM5000C and is the answer not in the manual?

Fazzini FM5000C Specifications

General IconGeneral
BrandFazzini
ModelFM5000C
CategoryMonitor
LanguageEnglish