13-1
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
Chapter 13 Maintenance
13.1 Maintenance Check
The overall check of the monitor, including a safety check, should be performed only by qualified personnel
before first use, every 6 to 12 month, and each time after repair.
Before using the monitor, do the following:
(a) Check if the work environment and power supply meet the requirements.
(b) Check if there is any mechanical damage.
(c) Check if the cables are worn and ensure insulation is in good condition.
(d) Check all the functions of the monitor to make sure that the monitor is in good condition.
(e) Check if the accessories used are specified by the manufacturer.
(f) Check the battery.
(g) If the monitor is equipped with a recorder, please check if the recorder is normal and recording paper
meets the specified specification.
(h) Check if the wiring resistance and leakage current meet the requirements.
If you find any damage on the monitor, stop using the monitor on patient,and contact the biomedical
engineer of the hospital or our customer service immediately.
All the safety and maintenance checks that need to dismantle the monitor should be performed by a qualified
customer service technician. Non-professional operation can cause the monitor damage or cause a security risk,
and human health may be endangered.
The circuit diagrams of the monitor can be provided by our company as per customer demands. Qualified
technicians can use it to help the user repair some apparatus that Our company classifies as can be maintained
by the user”.
Warning
If the hospital or agency that is responsible for using the monitor does not follow a satisfactory
maintenance schedule, the monitor may become damaged, and human health may be
endangered.
13.2 Maintenance Schedule
The following safety and maintenance check can be conducted by professional persons approved by Our
company. You can contact the customer service technicians if you need the following maintenance. Before
the inspection or maintenance, the facilities should be cleaned and disinfected.