FM5000C - FM5000PLUS
Maternal/Fetal Monitor
5.3 Module Setup
1. Access [MAIN MENU]→[MONITOR SETUP]→[MODULE SETUP];
2. Click the a certain parameter module to enable or disable its measurement function;
The fetal module defaults to on under fetal monitoring screen and mother/fetus monitoring screen and cannot be
disabled. While the fetal module defaults to off under mothering screen, trend screen and oxyCRG screen and
cannot be enabled.
5.4 Monitor Information
Monitor information is mainly for viewing the software version information operated by the monitor, so as to
facilitate the manufacturer to maintain and trace the monitor.
Access [MAIN MENU] and click [INFO] to view the software version information.
5.5 Demo Function
To enable Demo function:
1. Access [MAIN MENU];
2. Click [DEMO] and enter the password
Warning
Demo waveform is a kind of simulation of waveforms, which is made by the manufacturer only
to demonstrate the machine performanceand help users to conduct training.In actual clinical use,
the demo function should be disabled, because the medical staff may mistake it as the monitoring
waveforms and parameters of the patient, which affects monitoring and delays
diagnosis and treatment.
5.6 Locking the Touch Screen
To lock the touch screen:
Press on-screen key and hold for more than three seconds until the turns to , which
indicates the touch screen is locked.
To unlock the touch screen:
Press and hold for more than three seconds until it turns
to
, which indicates the touch
screen is unlocked.