EasyManua.ls Logo

Fazzini FM5000C - Page 116

Fazzini FM5000C
162 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18
-
3
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
Figure 18-2 Cuff usag
e
Attention
The width of the cuff should be 40 (or 50% for neonates) of the circumference of the limb,
or 2/3 of the length of the upper arm. The length of the inflated part of the cuff should be long
enough to wrap 5080% of the limb; a cuff of wrong size will produce a wrong reading. If you
have a problem about the size of the cuff, use a larger cuff in order to reduce errors.
Adult/pediatric/neonateresuable
cuff:
Patient type
Limb Circumference
Cuff width
Gastube length
Neonate
611cm
5cm
2 m
Infant
1019 cm
8 cm
Pediatric
1826 cm
10.6 cm
Adult 1
2535 cm
14 cm
Adult2
3347 cm
17 cm
Leg
4666 cm
21 cm
Neonate disposable cuff:
Size
Limb Circumference
Cuff width
Gastube length
Neonate 1
3.0 5.5 cm
2.6
cm
2 m
Neonate 2
4.0 7.6 cm
3.7 cm
Neonate 3
5.6 10.6 cm
4.5 cm
Neonate 4
7.0 12.8 cm
5.3 cm
Neonate 5
8.9 15.0 cm
6.0 cm

Table of Contents