FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
(11) Pulse Rate
Item
Specification
Measurement
range and accuracy
a) Analog SpO
2
module
Measuring range: 25bpm~250bpm;
Resolution: 1bpm;
Accuracy: ±1bpm.
b) Masimo SpO
2
module
Measurement range: 25bpm~240bpm;
Resolution: 1bpm;
Accuracy: ±3bpm (with motion) or ±5bpm (without motion).
c) Nellcor SpO
2
module
Measurement range: 25bpm~250bpm;
Resolution: 1bpm;
Accuracy: ±3bpm within the measurement range of 20bpm250bpm;
Not specified within the measurement range of 251bom300bpm.
Alarm range and
accuracy
Alarm range: 20 bpm300 bpm
Accuracy: ±1bpm
(12) Body Temperature
Item
Specification
Measurement
range
0~50
Accuracy
±0.2(sensor included)
Alarm range
0℃~50.0(32℉~122)
Alarm resolution
±0.1(±0.2)
Display resolution
0.1C
Operating mode
Derect mode
Transient response
No less than 20 seconds