III-6
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
TEMP COMM STOP
High
communicate with the
system.
T1 SENSOR OFF
Low
TEMP probe has not
connected well.
Check the connection of
TEMP probe.
Low power level! Please
charge the battery
Medium
Low battery
Please charge the battery
immediately.
CNNT PRS PRB
Low
The pressure transducer has
not connected.
Check the connection of
pressure transducer.
CNNT 1ST FHR PRB
Low
FHR1 transducer has not
connected.
Check the connection of
FHR1 transducer.
CNNT 2ND FHR PRB
Low
FHR2 transducer has not
connected.
Check the connection of
FHR2 transducer.
The first fetal wireless
probes battery is low.
High
The battery level of
FHR1 wireless transducer is
too low.
Change the battery of
FHR1 wireless transducer.
The second fetal wireless
probes battery is low.
High
The battery level of
FHR2 wireless transducer is
too low.
Change the battery of
FHR2 wireless transducer.
Uterine
contractions wireless
probes battery is low.
High
The battery level of
TOCO wireless transducer
is too low.
Change the battery of
TOCO wireless transducer.
ADJT ULTRAS
POSITION
High
FHR is the same as maternal
heart rate.
Please adjust the position of
FHR transducer.
RECORDER PAPER
LACK
High
No record paper loaded.
Load record paper.
Paper Tray Not Closed
High
The paper tray has not
closed well.
Close the paper tray well.
MONITOR STOP
High
Timed monitoring stops.
TIMED MONI FINISH
High
Timed monitoring ends.
(3) Prompt Message
Prompt message
Causes
LOOSE CUFF
NIBP cuff is too loose or has not connected with the monitor.
AIR LEAK
There is a leak in NIBP airway.
PNEUMATIC LEAK
AIR PRESSURE ERROR
An error occurred to the air pressure when measuring NIBP.
WEAK SIGNAL
The patients pulse is too weak or the cuff is too loose when measuring NIBP.
RANGE EXCEEDED
The NIBP measured value is not within the specified measurement range.
EXCESSIVE MOTION
The patients arm moves too frequently.
OVER PRESSURE
NIBP airway is occluded.
SIGNAL SATUATED
The NIBP signal is saturated due to excess motion or other sources.
NIBP SYSTEM FAILURE
An error occurred during NIBP measurement and the monitor cannot perform
analysis correctly.
NIBP TIME OUT