FM5000C - FM5000PLUS
Maternal/Fetal Monitor
communicate with the
system.
TEMP probe has not
connected well.
Check the connection of
TEMP probe.
Low power level! Please
charge the battery
Please charge the battery
immediately.
The pressure transducer has
not connected.
Check the connection of
pressure transducer.
FHR1 transducer has not
connected.
Check the connection of
FHR1 transducer.
FHR2 transducer has not
connected.
Check the connection of
FHR2 transducer.
The first fetal wireless
probes battery is low.
The battery level of
FHR1 wireless transducer is
too low.
Change the battery of
FHR1 wireless transducer.
The second fetal wireless
probes battery is low.
The battery level of
FHR2 wireless transducer is
too low.
Change the battery of
FHR2 wireless transducer.
Uterine
contractions wireless
probes battery is low.
The battery level of
TOCO wireless transducer
is too low.
Change the battery of
TOCO wireless transducer.
FHR is the same as maternal
heart rate.
Please adjust the position of
FHR transducer.
The paper tray has not
closed well.
Close the paper tray well.
(3) Prompt Message
NIBP cuff is too loose or has not connected with the monitor.
There is a leak in NIBP airway.
An error occurred to the air pressure when measuring NIBP.
The patient’s pulse is too weak or the cuff is too loose when measuring NIBP.
The NIBP measured value is not within the specified measurement range.
The patient’s arm moves too frequently.
The NIBP signal is saturated due to excess motion or other sources.
An error occurred during NIBP measurement and the monitor cannot perform
analysis correctly.