EasyManuals Logo

Fazzini FM5000C User Manual

Fazzini FM5000C
162 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
11-5
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
following tables are instructions of the four criteria.
NST score criterion is as follows:
Items
0 point
1-point
2-point
Results
Score
FHR baselinebpm
100,180
100109,161180
110160
FHR upgoing time
during FMs
09
1014
14
FHR upgoing range
during FMbpm
09
1014
14
Variation range
bpm
04
59,30
1030
FMbpm
None
102
2
KREBS score criterion is as follows:
Items
0 point
1-point
2-point
Results
Score
FHR baselinebpm
100,180
100109,161180
110160
Variation range
bpm
04
59,25
1025
Variation frequency
cpm
02
36
6
Accelerate
times/20mins
None
14
4
Decelerate
times/20mins
2
12
None
FMbpm
None
14
4
Advanced Fischer score criterion is as follows:
Items
0 point
1-point
2-point
Results
Score
FHR baselinebpm
100,180
100109,161180
110160
Variation range
bpm
04
59,30
1030
Variation frequency
cpm
01
26
6
Accelerate
times/20mins
None
14
4
Decelerate
times/20mins
Decelerate
later
or severe
variation
Slight variation
None
Fischer score criterion is as follows:
Items
0 point
1-point
2-point
Results
Score
FHR baselinebpm
100,180
100109,161180
110160
Variation range
bpm
04
59,25
1025
Variation frequency
bpm
01
25
5
Accelerate
times/20mins
None
Periodic acceleration
Non-Periodic
acceleration
Decelerate
times/20mins
Decelerate
later
or severe
variation
Slight variation
none

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fazzini FM5000C and is the answer not in the manual?

Fazzini FM5000C Specifications

General IconGeneral
BrandFazzini
ModelFM5000C
CategoryMonitor
LanguageEnglish