15  Data transmission
15.4  Data format FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
108
15.4 Data format
• Select the menu item Special Funct.\SYSTEM settings\serial transmis.
• Press ENTER until SER:kill spaces is displayed.
• Select on if the space characters are not to be transmitted.
• Press ENTER.
The file size will be considerably reduced (shorter transmission time).
• Select the decimal marker to be used for floating-point numbers (point or comma).
• Press ENTER.
This setting depends on the setting of the operating system of the PC.
• Select the character to be used to separate columns (semicolon or tabulator).
• Press ENTER.
15.5 Data structure
First, the header is transmitted. The first 4 lines contain general information about the transmitter and the measurement.
The following lines contain the parameters for each channel.
Next, the  line  \DATA is transmitted, followed by the column titles (see Tab.  15.2) for the corresponding channel. The
measured values are transmitted afterwards.
Special Funct.\SYSTEM settings\serial transmis.\SER:kill spaces
Special Funct.\SYSTEM settings\serial transmis.\SER:decimalpoint
Special Funct.\SYSTEM settings\serial transmis.\SER:col-separat.
Example
\DEVICE
\MODE
DATE
TIME
Par.Record
Meas.Point No.:
Pipe
 Outer Diameter
 Wall Thickness
 Roughness
 Pipe Material
 Lining
Medium
 Medium Temperat.
 Fluid pressure
Transducer Type
Sound Path
Transd. Distance
Damping
Full-Scale Val.
Physic. Quant.
Unit Of Measure
Numb.Of Meas.Val
: F50x -XXXXXXXX
: ONLINE
: 2018-01-09
: 19:56:52
: A:F5050
: 60.3 mm
: 5.5 mm
: 0.1 mm
: Carbon Steel
: WITHOUT LINING
: Water
: 38 C
: 1.00 bar
: xxx
: 3 NUM
: -15.6 mm
: 20 s
: 4.50 m3/h
: Volume flow
: [m3/h]/[m3]
: 100