16  Advanced functions
FLUXUS F502TE  16.2  Upper limit of the flow velocity
111
UMFLUXUS_F502TEV1-5EN, 2020-05-29
Totalizer behavior after the measurement is stopped
The totalizer behavior when the measurement is stopped or after a reset of the transmitter is set in the menu item 
Special Funct.\SYSTEM settings\Measuring\Quantity recall.
• If on is selected, the values of the flow totalizers will be stored and used for the next measurement.
• If off is selected, the flow totalizers will be reset to zero.
• Press ENTER.
Totalizer overflow
The overflow behavior of the totalizers can be set:
• Select the menu item Special Funct.\SYSTEM settings\Measuring.
• Select on to work with overflow.
The flow totalizer will be reset to zero automatically when ±9999999999 is reached.
• Select off to work without overflow.
The value of the totalizer increases to the  internal limit of 10
38
. The values will  be displayed as  exponential numbers
(±1.00000E10), if necessary. The flow totalizer can only be reset to zero manually.
• Press ENTER.
Independent of the setting, the flow totalizers can be reset to zero manually.
16.2 Upper limit of the flow velocity
Single outliers caused by heavily disturbed surroundings can appear among the measured values of the flow velocity. If
these outliers are not ignored, they will affect all derived physical quantities, which will be unsuitable for the integration
(e.g., pulse outputs).
It  is  possible  to  ignore  all measured  flow  velocities  exceeding  the  preset  upper  limit.  These measured  values will  be
marked as outliers.
The upper limit of the flow velocity is set in the menu item Special Funct.\SYSTEM settings\Measuring\
Velocity limit.
• Enter zero to switch off the outliers check.
• Enter a limit > 0 to switch on the outliers check. The measured flow velocity will then be compared to the entered upper limit.
• Press ENTER.
If the flow velocity is higher than the upper limit,
• the measured value will be marked as invalid. The physical quantity cannot be determined.
• an exclamation point (!) will be displayed after the unit of measurement (in case of a normal error, an interrogation point 
(?) is displayed).
Special Funct.\SYSTEM settings\Measuring\Quantity recall
Special Funct.\SYSTEM settings\Measuring\Quant. wrapping
Notice!
The overflow of a totalizer influences all output channels, e.g., the data logger and the online transmission of data.
The output sum of both totalizers (throughput Q) transmitted via an output is not valid after the first overflow (wrapping) 
of one of the totalizers.
In order to signalize the overflow of a totalizer, an alarm output with the switching condition QUANT. and the type 
HOLD has to be activated.
Special Funct.\SYSTEM settings\Measuring\Velocity limit
Notice!
If the upper limit is too low, a measurement might be impossible because most of the measured values will be marked 
as invalid.