9 Measurement
9.1 Parameter input
The pipe and fluid parameters are entered for the selected
measuring point. The parameter ranges are limited by the
technical characteristics of the transducers and the trans-
mitter.
• Select the program branch Parameter.
• Press ENTER.
9.1.1 Input of pipe parameters
Outer pipe diameter/pipe circumference
• Enter the outer pipe diameter. 
• Press ENTER.
An error message will be displayed if the entered parameter is outside the range. The limit is displayed.
Example: upper limit 1100 mm for the connected transducers and for a pipe wall thickness of 50 mm
Warning!
Installation, connection and start-up by unauthorized and unqualified personnel
This may result in personal or material damage or dangerous situations.
→ Any work on the transmitter has to be carried out by authorized and qualified personnel.
Caution!
Safety and accident prevention regulations for electrical systems and equipment
Failure to observe these regulations may lead to severe injuries.
→ Observe the safety and accident prevention regulations for electrical systems and equipment.
Notice!
The parameters will not be stored before the program branch Parameter has been edited in its entirety.
Parameter
Outer Diameter
Wall Thickness
Pipe Material
Lining
Roughness
Medium
Medium Temperat.
Transducer Type
Additional cable
see annex A, p. 146
Parameter\Outer Diameter
Outer Diameter
1100.0  MAXIMAL