17 SuperUser mode
FLUXUS F502TE 17.10 Display of the last valid measured value
127
UMFLUXUS_F502TEV1-5EN, 2020-05-29
17.10 Display of the last valid measured value
If the signal is not sufficient for a measurement, UNDEF is normally displayed. Instead of UNDEF, it is also possible to display
the last valid measured value.
Select the menu item Special Funct.\SYSTEM settings\Measuring\Miscellaneous.
Press ENTER until the menu item Keep display val is displayed.
Select on to activate the display of the last valid measured value or off to deactivate it.
Press ENTER.
17.11 Displays during the measurement
Besides the normal information (see section 9.4) the following parameters can be displayed during the measurement in
the SuperUser mode:
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Keep display val
display explanation
t transit time of the measuring signal in the fluid
c sound speed
REYNOLD Reynolds number
VARI A standard deviation of the signal amplitude
VARI T standard deviation of the transit time of the measuring signal
dt-norm transit time difference standardized to the transducer frequency
fluid density