19 Thermal energy rate measurement
19.2 Transmitter settings FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
138
19.2 Transmitter settings
If the position of the transducers and the temperature inputs are changed, the following settings have to be made.
19.2.1 Assignment of the transducers and the temperature inputs
The position of the transducers and the temperature inputs can be assigned in accordance with the application.
• Select the menu item Special Funct.\SYSTEM settings\Proc. inputs\Link temperature.
• Press ENTER.
• Select heat for a heating application or chill for a cooling application.
• Press ENTER.
• Select return if the transducers are mounted on the return line or supply if they are mounted on the supply line.
• Press ENTER.
• Select sign if the sign of the thermal energy rate is to be considered or absolute if just the absolute value of the thermal
energy rate is to be displayed.
• Press ENTER.
• Select yes if the aggregate state of the fluid changes between the return and supply line. Select no if it does not change.
• Press ENTER.
• Select the temperature input to be assigned to the supply temperature.
• Press ENTER.
• Select the temperature input to be assigned to the return temperature.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. inputs\Link temperature
Special Funct.\...\Thermal energy
Special Funct.\...\Transd. Location
Special Funct.\...\Thermal energy
Special Funct.\...\With PhaseChange
Special Funct.\...\T-Supply\Input T1
Special Funct.\...\T-Return\Input T2