6  Mounting
6.2  Transducers FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
28
6.2.2 Installation of the transducers
6.2.2.1 Orientation of the transducers and determination of the transducer distance
Observe the orientation of the transducers. If the transducers have been mounted properly, the engravings on them form
an arrow (see Fig. 6.5). The transducer cables show in opposite directions.
The transducer distance is measured between the inner edges of the transducers.
• Select the installation instructions that correspond to the supplied transducer clamping fixture.
Bidirectional measurements
• The sound beam plane is selected according to the 
nearest elbow (horizontal or vertical, depending on the 
pipe orientation, see above).
Fig. 6.5: Orientation of the transducers and transducer distance
a – transducer distance
flow direction
a