6  Mounting
FLUXUS F502TE  6.2  Transducers
41
UMFLUXUS_F502TEV1-5EN, 2020-05-29
6.2.2.5 Mounting with tension strap and quick release clasp
The clasp is fixed to the tension strap (see Fig. 6.33).
• Cut the tension strap to length (pipe circumference + 120 mm).
• Position the sensor on the pipe (see Fig. 6.34).
• Place the tension strap around the transducer and the pipe.
• Place the tension strap around the pipe and insert it into the clasp.
• The clasp has to seat solidly on the pipe.
• Tighten the tension strap.
• Tighten the clasp screw.
Caution!
The edge of the tension strap is very sharp.
Risk of injury! 
→ Debur sharp edges. 
→ Wear the required personal protective equipment. 
→ Observe the applicable rules. 
Fig. 6.33: Tension strap with quick release clasp
1 – clasp screw
Fig. 6.34: Transducer on the pipe
1