7 Connection
7.1 Transducers FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
48
• Remove the blind plug for the connection of the transducer cable.
• Insert the transducer cable into the housing.
• Fix the transducer cable by tightening the cable gland.
• Connect the transducer cable to the terminals of the transmitter (see Fig. 7.3 and Tab. 7.2).
Fig. 7.3: Connection of the transducer cable with plastic jacket and stripped ends to the transmitter
1 – cap nut
2 – compression part
3 – basic part
4 – external shield, brushed back
Tab. 7.2: Terminal assignment
terminal connection
AV transducer (brown cable, marked white)
AVS transducer (red cable)
ARS transducer (red cable)
AR transducer (brown cable)
70 mm
20 mm
10 mm
4
cable gland
1 2 3