EasyManua.ls Logo

Fluid Components Intl ST51A - Z Menu: Configure Flow Units and 4-20 Ma Output Scaling; Figure 25 - Z Menu Command Structure: Units and Scaling Setup; Table 6 - Typical Serial Interface Top Level Commands for Flow Meter Configuration

Fluid Components Intl ST51A
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ST51A/ST75A/ST75AV Mass Flow Meters OPERATION
Fluid Components International LLC 29
Table 6 – Typical Serial Interface Top Level Commands for Flow Meter Configuration
Command
Name
Description
T
Normal Operation Mode
Normal operating mode: All outputs are active.
Z
Flow Unit Set-up
Set English/Metric flow units; set up pipe dimensions for volumetric units.
V
Output Configuration
Set 4-20 mA and pulse output configuration.
S
Totalizer Menu
Enables W menu (option)
W
Totalizer
Enable/Disable
F
K-Factor (default=1)
Flow factor
N
Warm Reset
Reinitialize Instrument
Z Menu: Configure Flow Units and 4-20 mA Output Scaling
Use the Z menu to change flow units. Note, however, that changing units requires rescaling of the unit (set new zero and span). The 4-20 mA Zero
and Span can be changed from the original calibration, provided the new values are within the original calibrated range; i.e., if the original
calibration was 1 to 100 SCFM (4-20 mA), the new zero (4 mA) must be equal to or greater than 1 SCFM and the new span (20 mA) must be
equal to or less than 100 SCFM. The flowchart in Figure 25 below gives an overview of the instrument’s Z menu programming.
Note:
The Z menu is passcode protected when the totalizer function is enabled. Contact the factory for details.
Figure 25Z Menu Command Structure: Units and Scaling Setup

Table of Contents

Related product manuals