OPERATION ST51A/ST75A/ST75AV Mass Flow Meters
50 Fluid Components International LLC
Command 146: Write Customer Flow Units
Byte
Format
Description
Request Data Bytes
0
Unsigned-8
Units Code for Flow
Response Data Bytes
0
Unsigned-8
Units Code for Flow
Response Codes:
See Table 16, page 53, for response code list.
Command 148: Read Plenum Information (Pipe Size)
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0-3
Float
Pipe Height Value
4–7
Float
Pipe Width (Diameter) Value
8
Unsigned-8
Plenum Units Code
Response Codes:
See Table 16, page 53, for response code list.
Command 149: Write Plenum Information (Pipe Size)
Byte
Format
Description
Request Data Bytes
0–3
Float
Pipe Height Value
4–7
Float
Pipe Width (Diameter) Value
8
Unsigned-8
Plenum Units Code
Response Data Bytes
0-3
Float
Pipe Height Value
4–7
Float
Pipe Width (Diameter) Value
8
Unsigned-8
Plenum Units Code
Response Codes:
See Table 16, page 53, for response code list.
Command 151: Read Calibration Flow Limits
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–3
Float
Flow Lower Limit Value
4–7
Float
Flow Upper Limit Value
Response Codes:
See Table 16, page 53, for response code list.
Command 152: Write Calibration Flow Limits
Byte
Format
Description
Request Data Bytes
0–3
Float
Flow Lower Limit Value
4–7
Float
Flow Upper Limit Value
Response Data Bytes
0–3
Float
Flow Lower Limit Value
4–7
Float
Flow Upper Limit Value
Response Codes:
See Table 16, page 53, for response code list.
Command 153: Read Calibration Temperature Limits
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–3
Float
Temperature Lower Limit Value
4–7
Float
Temperature Upper Limit Value
Response Codes:
See Table 16, page 53, for response code list.
Command 154: Write Calibration Temperature Limits
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–3
Float
Temperature Lower Limit Value
4–7
Float
Temperature Upper Limit Value
Response Codes:
See Table 16, page 53, for response code list.