5.3-12 Section 5: Using the Model 9100: Procedure Mode
Final Width = 215mm
5.3.3.12 User Options Following 'ABORT' or 'END'
Once the procedure has ended or been aborted, the user can return to one of five points
in the sequence. The point numbers refer to Fig 5.3.1:
USER (Point 1)
The 'Select USER NAME' screen is the first to appear after selecting Procedure mode.
All setup parameter variables can be changed, and a new procedure card can be inserted.
MODEL (Point 2)
The 'Select MODEL' screen offers user-selection from all the UUT models (for the
previously-selected manufacturer) on the currently-loaded procedure card; or a new
procedure card can be inserted.
Note: For those cards containing procedures for UUTs from more than one manufacturer:
then in order to change manufacturer, return to the 'Select USER NAME' screen
by pressing the 'USER' screen key.
SERIAL (Point 3)
This choice assumes a wish to select a different unit of the same model. The system
therefore returns to the 'Enter SERIAL NUMBER' screen.
PROC (Point 4)
This choice assumes a wish to select a different procedure for the same model and serial
number. The system therefore returns to the 'Select PROCEDURE' screen.
RETRY (Point 5)
This selection re-runs the same procedure which has just ended or been aborted. It is
assumed that the same unit is being tested, so the same serial number will appear on any
results print-out.