Section 3: Model 9100 Controls: Modes of Operation 3-23
Final Width = 215mm
TODAYS DATE TIME
TODAYS DATE TIME
3.3.2.21 'LAB TEMP'
1. On the LAB DETAILS 'Present
Settings' menu screen, press the
LAB TEMP screen key on the
right. This transfers to 'The
current temperature setting'
screen
2. Use the alpha-numeric keypad to
enter the laboratory temperature:
(2 characters and );
then the temperature tolerance:
(2 characters and ).
Configuration
EXIT
The current temperature setting is
21°C ±10°C
Enter the temperature °C
Laboratory Temperature
Users should enter the official
temperature of their laboratory for the
results card and the printed certificates.
[Backspace (
) deletes the entered
characters one at a time.
CLR (
then ) will return to the
LAB DETAILS screen.]
Completed entries will have been transferred to
the 'Present Settings' list; the parameters of
incompleted entries will remain unchanged.
3. To revert to the LAB DETAILS screen,
press the EXIT key.
Laboratory Humidity
Users should enter the official humidity
of their laboratory for results cards and
printed certificates.
3.3.2.22 'LAB HUMID'
1. On the LAB DETAILS 'Present
Settings' menu screen, press the
LAB HUMID screen key on the
right. This transfers to 'The
current humidity setting' screen
Configuration
EXIT
The current humidity setting is
30% ±10%
Enter the humidity %
2. Use the alpha-numeric keypad to
enter the laboratory humidity:
(2 characters and );
then the humidity tolerance:
(2 characters and )
Backspace (
) deletes the entered
characters one at a time.
CLR (
then ) will return to the
LAB DETAILS screen.
Completed entries will have been transferred to
the 'Present Settings' list; the parameters of
incompleted entries will remain unchanged.
3. Press the EXIT key to revert to the LAB
DETAILS screen.