Technical Data
Electrical Specifications
23
2
2.3 Electrical Specifications
Power 20 to 48 VDC, max. 12 W
Power over Ethernet (IEEE 802.3af, pinout mode A, mixed DC & data)
Outputs
Analog 0 to 20 mA (active), or
4 to 20 mA (active)
current loop impedance: max. 500 Ω
short circuit resistant
electrically isolated
Alarm 48 V / 300 mA
1 potential-free relay output with wear-free contacts (solid state relay),
electrically isolated
Inputs
Analog 0 to 20 mA or 4 to 20 mA
loop impedance max. 220 Ω
electrically isolated
emissivity setting, slope setting or
background temperature compensation
Digital trigger input (closing contact, active with low status)
electrically isolated
signal processing control
sighting control on/off
RS485
2-wire communications (half duplex)
networkable to 32 sensors
baud rate: 1200, 2400, 9600, 19200, 38400, 57600, 115200 Bit/s
(default: 38400 Bit/s)
data format: 8 bit, no parity, 1 stop bit
electrically isolated
Ethernet
Connection M12 socket, 4 wires (full duplex)
100 MBit/s, 100BASE-TX / IEEE 802.3u, Auto-Negotiation
electrically isolated
Addressing DHCP or fixed IP address
Defaults IP address: 192.168.42.132
subnet mask: 255.255.255.0
gateway: 192.168.42.1
Protocols UDP, port 36363 (fixed)
TCP/IP version 4, default port 6363, range: 1 to 65535
http web server (port 80)
MJPEG-Video
PROFINET IO, see section 9 PROFINET IO, page 67
EthernNet/IP, see section 10 EthernNet/IP, page 71.