6.3.2.3 Signal Drop with Reset
The following three signal drop (decay) functionalities are implemented:
Signal Drop: Perpendicular
The perpendicular signal drop (default mode) is activated, if both relevant signal decay values (linear & averaging
drop) are set to zero (0.0 Kelvin / s).
Figure 6-5: Signal Drop: Perpendicular
Signal Drop: Linear
The signal drop follows a linear decay function, where the decay value is given in Kelvin / s.
ASCII command: <XE>
Figure 6-6: Signal Drop: Linear
Signal Drop: Averaging
The signal drop follows an averaging function. The average time is the amount of time the output signal needs to
reach 90% magnitude compared to a perpendicular drop.
ASCII command: <AA>
output temperature
object temperature
Temp
Time
output temperature
object temperature
Temp
Time