EasyManua.ls Logo

Fluke endurance series - Environmental Specifications

Fluke endurance series
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Endurance
®
Series
Users Manual, Rev. 2.0, Jun 2020
24
2.4 Environmental Specifications
Ingress Protection IP65 (IEC529) / NEMA 4 (with end cap screwed on)
Operating ambient temperature
Without cooling:
E2RL 0 to 60°C (32 to 140°F)
all other models 0 to 65°C (32 to 149°F)
With air cooling:
all models 0 to 120°C (32 to 248°F)
With water cooling:
all models 0 to 175°C (32 to 347°F)
Water cooled by ThermoJacket:
all models 0 to 315°C (32 to 600°F)
Internal case temperature
all models 10 to 72°C (50 to 162°F)
Storage temperature -20 to 70°C (-4 to 158°F)
Humidity 10% to 95% @ 22 to 43°C (72 to 109°F), non-condensing
Shock IEC 68-2-27 (5 G, 11 ms duration, 3 axes)
Vibration IEC 68-2-6 (2 G, 10 to 150 Hz, 3 axes)
EMC EN 61326-1:2013 industrial
Warm up Period 15 min.
Material
Housing Stainless steel (Mat.-No.: 1.4305, X8CrNiS18-9)
Weight
Sensor 1220 g (43 oz)
Mounting nut 62 g (2.2 oz)
Fixed mounting bracket 264 g (9.3 oz)
Air/Water cooled housing 1760 g (62 oz)

Table of Contents