Operation
Menu Structure
47
6
6.2.2 Configuration Menu
CONFIGURATION
(1-color mode)
CONFIGURATION
(2-color mode)
MODE
2-COLOR
TEMP. UNITS
°C
TEMP. UNITS
°C
RELAY MODE
NORMALLY OPEN
RELAY MODE
NORMALLY OPEN
ATTENUATION
RELAY 95%
ATTENUATION
FAILSAFE 95%
LASER / LED / CAMERA
ON
LASER / LED / CAMERA
ON
FACTORY DEFAULT
NO
FACTORY DEFAULT
NO
KEY -ENTER-
UNLOCKED
KEY -ENTER-
UNLOCKED
MODE
The subentry is available for 2-color sensors only. It allows to force the device to operate either in 1-color or in 2-
color mode.
TEMP. UNITS
toggles between the temperature units °C or °F.
RELAY MODE
toggles between the relay modes for the potential free relay contact:
NORMALLY OPEN
PERMANENT CLOSED
PERMANENTLY OPEN
NORMALLY CLOSED
ATTENUATION RELAY
allows to set the attenuation threshold to switch the relay. Range is from 0 to 95% of attenuation.