18
2 Technical Data
2.1 Measurement Specification
Temperature Range
E1ML 400 to 1740°C (752 to 3164°F)
E1MH 540 to 3000°C (1004 to 5432°F)
E1RL 550 to 1800°C (1022 to 3272°F) 1 color mode
600 to 1800°C (1112 to 3272°F) 2 color mode
E1RH 1000 to 3200°C (1832 to 5792°F)
E2ML 250 to 1100°C (482 to 2012°F)
E2MM 250 to 1400°C (482 to 2552°F)
E2MH 450 to 2250°C (842 to 4082°F)
E2RL 250 to 1200°C (482 to 2192°F)
E3ML 50 to 1000°C (122 to 1832°F)
E3MH 150 to 1800°C (302 to 3272°F)
Spectral Response
E1ML 1 µm
E1MH 1 µm
E1RL 1 µm
E1RH 1 µm
E2ML 1.6 µm
E2MM 1.6 µm
E2MH 1.6 µm
E2RL 1.6 µm
E3ML 2.4 µm
E3MH 2.4 µm
Response Time
1
E1ML 2 ms
E1MH 2 ms
E1RL 10 ms
E1RH 10 ms
E2ML 2 ms
E2MM 2 ms
E2MH 2 ms
E2RL 20 ms
E3ML 20 ms
E3MH 20 ms
1
95% value