If you are dispensing to deep well blocks, you will need pressurized bottles and
continuous flow chips. You will also need to calibrate the reagents for continuous
flow use.
● Insert the tubes into the holes in the CF chip.
● Tighten the screw in the CF chip by using a 1.5 mm screwdriver.
2. Put the chip pallets into the chip port stations.
3. Attach the reagent tubes or pipette tips to the ingredient input at each chip. Skip
this step if you are using CF chips.
● If you are using a pipette tip, you will need to remove the black adapter from
the chip’s input.
MANTIS User's Guide | Chapter 8: Setting Up the Hardware
Note: It is recommended to cut your input tubing to a length of 45 cm for
High Volume and Low Volume chips.
Installing a Continuous Flow Chip
Putting the Chip into the Chip Port