9.5 Adding columns to the column list
Adding columns
It is possible to add columns, for example “RTP columns on request”, to the ÄKTA ready
installation wizard. Follow the steps below to add a new column to the UNICORN column
list.
In UNICORN 7 it is possible to add columns and set basic physical parameters
in the Column Handling dialog box, see UNICORN Method Manual. Advanced
column parameters (e.g., PID parameters) must be added to the method where
the column is used, see Edit column parameters in UNICORN 7, on page 130 .
Note:
Instruction
ActionStep
Click Column List on the Edit menu in the Method Editor module.1
Select a column supported by the
ÄKTA ready software (i.e. a ReadyTo-
Process column with the prefix “RTP”
in the list), that has the same column
volume and average particle diame-
ter as the column to be added
(marked items in figure). This will set
the advanced parameters that are
compatible with the software.
2
Click Edit for the selected column.3
Enter appropriate parameter values for the column to be added (Max pres-
sure, for instance, depends on the type of medium).
4
Click Save As to save the column with
a new name. Save as global only if
the column is to be available to other
users (e.g., on a network).
5
Note:
Do not click Save. If a column is edited and you click Save, the edited param-
eters will be overwritten.