3.3 Flow chart
Chart and components
The liquid flow through ÄKTA ready is illustrated in the chart below.
(Black: standard components; Red: optional components)
XV-056
XV-055
XV-054
XV-053
XV-052
XV-051
XV-001
XV-002
XV-003
XV-004
XV-005
XV-006
PT-111
PT-112 PT-113
AE-121 TE-161 AE-131
FE-141 CE-101
AT-221
HV-301
P-201
A B C D
P-202
The table below specifies the maximum operating pressure the flow path could be
pressurized to with the pressure upgrade strategy in the different parts of the system.
PressureDescriptionPart
0.6 barLow pressure, over the Inlet ManifoldA
5.0 barHigh pressure, over the air trap manifold and columnB
2.0 barMedium pressureC
0.95 barMedium pressure, over the outlet manifoldD
The table below describes the system components shown in the flow chart.
FunctionPart
Inlet 1 to 6, grouped in Inlet ManifoldXV-001 to XV-006
ÄKTA ready Operating Instructions 28960345 AD 45
3 System description
3.3 Flow chart