5.2 Start the system
General
ÄKTA ready is normally left switched on between runs. The unit is switched off only during
monthly maintenance and repair, or when not in operation for a longer period of time
(more than a few days).
Starting ÄKTA ready
ActionStep
Switch on power to ÄKTA ready by turning the Mains Power switch to the
"I" position. See Components on system cabinet front, on page 38 for loca-
tion.
1
Make sure that UNICORN is started according to the instruction above. Wait
for connection.
2
After this sequence is completed, the following occurs:
The POWER indicator lamp flashes. See Components on system cabinet front, on
page 38 for location.
When communication with UNICORN software is established, the POWER indicator
lamp shows a steady green light.
When power is switched on ÄKTA ready will perform a stray light calibration
without the UV flow cell installed. This will result in a very high calibration
factor. Therefore it is very important to perform a stray light calibration as
soon as the flow kit is mounted and water or any other actual liquid has been
purged through the flow kit.
Note:
76 ÄKTA ready Operating Instructions 28960345 AD
5 Starting ÄKTA ready
5.2 Start the system