Corrective actionProblemComponent
Emergency button has been pressed.
Pump lid not properly attached / secured. See
Connect and mount the pump tubing, on page 89
for instructions.
No inlet or outlet valve open. Check method and
valves.
Incorrect method. Check by entering
Pump:ManFlow > 1%.
If none of the above, contact GE Healthcare service
personnel.
Pump not workingPump
Check that connected inlet is actually used.
Inlet containers are placed too low relative to the
pump. Check inlet containers.
No liquid is supplied to the pump. Check inlet con-
tainers.
Flow kit tubing is split or burst. Replace flow kit.
Kinks or blockages in the flow kit. Remount or re-
place flow kit. See Section 6.3 Mounting the flow kit,
on page 81 for mounting instructions.
Pump rotor slipping on the drive shaft, contact GE
Healthcare service personnel.
Little or no flow
Liquid too viscous and/or pump speed too slow. Check
that viscosity of liquid is within specified range (see
Section 14.1 Specifications, on page202). Increase pump
speed.
Pulsation of liquid in flow
kit
Check that connected outlet is actually used.
Check that the outlet is not plugged.
Too small outlet tubing diameter. Change to a
larger tubing.
Flow kit mounted on system for too long, causing
the inner walls of the tubing to stick together. Re-
place the flow kit.
Too high outlet pressure