WARNING
Never exceed the operating limits stated in this document and
on the system label. Operation of the product outside these
limits can damage equipment and cause personal injury or
death.
High pressure. The flow rate may under no circumstances
exceed the specified column maximum flow rate. High flows
might affect the packed medium, causing the pressure to ex-
ceed the specified column maximum pressure.
Use columns that withstand expected pressures. If not, the
columns might rupture, resulting in injury. Preferably use
ReadyToProcess columns from GE Healthcare.
Over-pressure. Never block the outlet tubing with, for instance,
stop plugs, since this will create over-pressure and might result
in injury.
Never touch the pump or pump lid while the pump is running.
Never put fingers or any objects other than the intended tubing
in to the pinch valve opening.
Corrosive substance. NaOH is corrosive and therefore danger-
ous to health. When using hazardous chemicals, avoid spillage
and wear protective glasses and other suitable Personal Pro-
tective Equipment (PPE).
Emergency stop. Pressing the EMERGENCY STOP will not shut
off mains power to the cabinet.
Power failure. During a power failure, or if the EMERGENCY
STOP button is pressed, the equipment may remain pressur-
ized. Opening a line or vessel at this point could result in the
release of potentially hazardous process or cleaning fluid, and
cause bodily harm.
When recovering from a power failure or emergency shutdown,
make sure all lines and vessels are depressurized before
opening.
Emergency stop. Pressing the EMERGENCY STOP does not
automatically depressurize the flow path.
20 ÄKTA ready Operating Instructions 28960345 AD
2 Safety instructions
2.1 Safety precautions