Before starting
System setup
In UNICORN 5 the component list is reached by clicking System Setup on the Admin-
istration menu in the UNICORN Manager
In UNICORN 7, in the Administration module, click System Properties and then click
Edit in the System Properties dialog box.
Select the Profibus_Config5 or Profibus_Config6 checkboxes. If you have installed the
gradient system both the Äkta_Ready_Gradient checkbox and the Profibus_Config6
checkbox must be selected.
You should also select the pH_Cell checkbox if pH is to be measured, the correct flow
kit size (the LowFlow_Kit checkbox, default, or the HighFlow_Kit checkbox), type of air
sensor (the Air_AE151 or the Air_AE152 checkbox). By selecting the Service checkbox
you will have the possibility to display UV_131_S and UV_131_R in RunData.
TC clamps
Two TC clamps are provided with the flow kit. An additional three clamps are required
in order to mount the flow kit and run a component test.
Aseptic solution
A suitable disinfectant solution should be available for aseptic treatment of tubing con-
nectors. Connectors can, for instance, be treated with 70% Ethanol before connecting.
82 ÄKTA ready Operating Instructions 28960345 AD