Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 6-9
adjacent to the end point.
When calibration is changed during the procedure, the distance measurements are recalculated and
display the new values.
To perform distance measurements
1. Perform Calibration.
2. Select the Distance button. A default measurement line is displayed. This line is based on the
calibration performed.
3. Select the distance measurements unit, it includes mm, inch and French.
4. Select the line and drag the anchor to the appropriate position. The distance measurement will display
adjacent to the anchor.
5. You can enter up to five distances per image.
To delete a Distance: select it on the right monitor and press Delete button on the screen.
To save a copy of the image with distance: press the Save key on the Workstation
membrane
keyboard.