Start Up and Operating Control
Page 4-16
WARNING
4.3.3. Emergency Switch
The emergency switch located on the top of C-Arm control panel. When emergency situation occurs, press
the switch to stop vertical column mechanical movement and disable X-Rays.
If pressing an emergency switch fails to stop motor movement or X-Rays, press the power
off switch or disconnect the power plug from the AC receptacle.
If the emergency switch is pressed on purpose or inadvertently, vertical column operation will cease and
X-Rays will be disabled. Rotate the Emergency Switch in the direction of the arrow indication to release it
and continue to use the system.
Should the system unexpectedly become unresponsive or lockup, restart the system to return it to normal
operating condition. If the system fails to respond or continues to lockup, call your service representative.