Maintenance
Page 9-10
9.5.2. Periodic Maintenance
Periodic Maintenance should be performed by a GE HUALUN medical systems Co. Ltd. representative or a
qualified service engineer.
The following periodic maintenance schedule is recommended.
Item
Description
Frequency
Earth (Ground)
Check maximum earth of whole system
Annually
Power supplies
Check earth leakage current
Annually
Check internally- generated DC voltages
Annually
UPS
Check no leakage
Annually
Check the appearance is normal
Annually
Check the cable connection
Annually
Check the UPS life, check the need for replacement
Annually
PCB’s and racks
Ensure secure fitting and check for dust and corrosion
Annually
Motorized movement
Check electrical and mechanical movement
Annually
Bearings
Check movement freedom
Annually
Controls and Indicators
Check accuracy and functioning of:
All controls
Annually
All visible / audible indicators
Annually
Movement
Annually
Dose / Dose rate / DAP
Annually
Alignment
Collimator alignment and field limitation
Annually
Beam alignment and centering
Annually
Mechanical
Counterweight fasteners and cables
Annually
All mechanical stops
Annually
Wheels and wheel alignment
Annually
Brakes and locks
Annually

Table of Contents

Other manuals for GE Brivo OEC 715

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Brivo OEC 715 and is the answer not in the manual?

GE Brivo OEC 715 Specifications

General IconGeneral
BrandGE
ModelBrivo OEC 715
CategoryMedical Equipment
LanguageEnglish

Related product manuals