1.5.7. Ground Fault
If the operating room has a ground fault alarm and the alarm is actuated:
Do not operate the system.
Call a qualified service technician.
1.5.8. Improper Access
Improper access into the cabinet and covers of the system may cause injury to your hand or
fingers. Do not put your hands or fingers into the cabinet or covers of the system.
1.6. Radiation Exposure
1.6.1. General Protection
This equipment either produces or is used in the vicinity of ionizing radiation. Observe
proper safety practices during operation.
The owner must designate areas suitable for safe operation and service of the equipment and ensure they
are only used in those areas.
The owner must ensure that all personnel wear appropriate protective clothing and radiation monitoring
devices while using the equipment.
Remain alert for visual indicators and audible alarms that are activated when ionizing radiation is being
produced by equipment in the work area.
1.6.2. Source-to-Skin Distance
International regulations specify that a minimum source-skin distance be maintained, except for specific
surgical applications. Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product has a skin spacer attached in
order to meet this requirement. The system is prefabricated with a spacer for 30cm source-to-skin
distance.
Operator should keep as far as possible away from the X-Ray source.
Removing the skin spacer may result in increased radiation exposure to the patient. The
skin spacer can only be removed by the physician.