Electrical circuits inside the equipment may use voltages that are capable of causing
serious injury or death from electric shock. To avoid this hazard, never remove any of the
system covers.
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product is not waterproof. If water, soap or
other liquids drip into the equipment, this can cause short circuit leading to electric shock
and fire hazards. If liquids are accidentally spilled into the system, do not power or turn on
the system until the liquids dry or evaporate completely.
To avoid risk of electric shock, this equipment must only be connected to a supply mains
with protective earth.
Observe the following safety procedures to avoid electric shock or serious injury to operators and patients
and to avoid system malfunction.
Make all electrical connections to equipment while outside the patient environment. Do not touch a
connector and the patient at the same time.
Do not bypass, jumper or otherwise disable the safety interlocks.
Do not remove any of the assembly covers.
Do not place food or beverage containers on any part of the equipment. If spilled they can cause short
circuits.
Always remove power to the equipment before cleaning. Use a slightly damp cloth or sponge for cleaning.
Only qualified service engineers are allowed to service or repair a system.
1.5.6. Electrical Fire
In the event of electrical fire perform the following emergency procedure:
Any emergency procedure developed by the owner, for the area in which the system is
used, should include these safety measures:
Remove electrical power to the system by placing the power switch in the off position.
Unplug the power cord from the AC receptacle.
Evacuate personnel from the area.
Only use a fire extinguisher that is approved for use on electrical fires.
Call your local fire department for help if necessary.
The use of the wrong type of fire extinguisher presents electrical shock and burn hazards.
To avoid these hazards, a fire extinguisher, which meets applicable regulations and
standards, must be available in the room where the equipment is being used.