Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 9-5
9.2.5. Film Mode Performance Check
This procedure produces X-Rays. Take the appropriate precautions.
1. Press the Film key on the C-Arm control panel to select film mode.
2. Install the film cassette holder on the image intensifier and load a film cassette.
3. Press the footswitch and handswitch and hold on until the exposure indicator light is off. The system
will give continuous audible beeps during the exposure. If the X-Ray switch is released early during
exposure, the system will give intermittent beeps to warn that the exposure is interrupted.
4. Develop the film and inspect the exposure.