Technical Reference
Page 12-2
12.1. Overview
The policy of GE Healthcare is one of continual product development and improvement.
For this reason, GE Healthcare reserves the right to change the operating characteristics and specifications
of newer products at any time, without prior notice, and without incurring any obligation relating to
previously manufactured items.
The specifications listed are limited to general performance and physical data. Specifications of optional
equipment provided by other manufacturers are given in the applicable manuals provided with those
options.
12.2. System Specifications
12.2.1. Classification Type
• Class I equipment (as defined by IEC60601-1)
• Type B protection against electric shock (IEC60601-1)
• Ordinary protection against ingress of water
Degree of Protection:
System: IPX0
Footswitch: IPX8
• Operation Mode:
Continuous operation
• Loading Mode:
Intermittent loading
• All parts of the equipment, including accessories and connecting cables, can be disinfected by wiping
them with a cloth dampened with disinfectant, such as 75% alcohol.
12.2.2. Environmental Requirements
Operating Conditions:
Temperature: +10°C~+40°C (+50F~+104F)
Humidity: 20%~80% Relative, non-condensing
Atmospheric Pressure: 700hPa~1060hPa
Transportation and Storage:
Temperature: -20°C~+55°C (-4F~+131F)
Humidity: 10%~95% Relative, non-condensing
Atmospheric Pressure: 500hPa~1060hPa